Infusion Devices - Syringe Pump: Baxter AS40A
The Baxter AS40A has advanced programming capability, including variable occlusion sensitivity, automatic syringe size recognition, and a Drug Library option with drug-specific infusion parameters stored in memory, and keypad lock. It can deliver continuously or can deliver a timed infusion of one or more equal doses according to a programmed schedule.
Dimensions: 3.4"W x 2.6"D x 10"H
Weight: 2.75 lbs
Power: AC: 105-125V Hz (battery charger); DC: internal nickel-cadmium battery pack |
 |
 |
OPERATING RANGE:
| Syringe Types: |
|
B-D Plastipak - 3, 5, 10, 20, 20, 60 mL
Monoject - 1, 3, 5, 12, 20, 35, 60 mL
Terumo - 1, 3, 5, 10, 20, 30, 60 mL
|
| Accuracy: |
|
±3% (not including syringe tolerance. For volume infusions 1.3%, or .007” of travel, whichever is greater (not including syringe tolerance) |
| Flow Rate Range: |
|
0.01 mL/hr to 360 mL/hr depending on syringe size |
| Deliverable Volume: |
|
Full syringe volume can be delivered |
| Data Display: |
|
Self-prompting, multi-field LCD |
| Status Display: |
|
Nine LED array |
| Battery Life: |
|
12 hours of operation at 2 mL/hr using a 60 mL syringe, following a 16 hour charge |
|
Our personnel are available 24/7 to handle most local rental requests. Please contact your local representative for more details.
Back to Previous Page
|
 |
 |
|